Specifications
Rosemount MS Magnetic Sensor MS030 for Slurry Applications, 3 inch (DN080) | |
---|---|
Reference Accuracy | ±0.25% with optional ±0.15% high accuracy |
Line Size | 3” (80mm) to 36” (900mm) |
Liner Material Options | PTFE, PFA+, Polyurethane, Neoprene, Linatex Natural Rubber, Adiprene |
Electrode Options |
Material: 316L Stainless Steel, Nickel Alloy, Platinum, Tantalum, Titanium Type: Standard, Bullet, Flat Optional 3rd electrode for grounding |
Flange Options |
Connections: Slip-on, weld-neck, grooved coupling Ratings: ASME Class 150 – 2500, EN-1092 PN10 – PN40, AS2129 Table D and E, JIS B2220 10K – 40K, AS4087 PN16 – PN35 Materials: Carbon Steel, 304/304L SST, 316/316L SST Additional flange ratings and materials may be available upon request |
Rosemount 8732E Field Mount Magnetic Flow Meter Transmitters | |
---|---|
Sensor compatibility | |
Compatible with Rosemount 8705, 8711, and 8721 sensors. Compatible with AC and DC powered sensors of other manufacturers. | |
Transmitter coil drive current | 500mA |
Flow rate range | Capable of processing signals from fluids that are traveling between 0.04 and 39 ft/s (0.01 to 12 m/s) for both forward and reverse flow in all sensor sizes. Full scale continuously adjustable between –39 and 39 ft/s (–12 to 12 m/s). |
Conductivity limits | Process liquid must have a conductivity of 5 microSiemens/cm (5 micromhos/cm) or greater. |
Power supply | 90 – 250VAC, 50/60Hz or 12 – 42VDC |
Line power fuses | 90-250VAC systems: 1A, 250V, I2t, 1.5 A2s Rating, Fast Acting Bussman AGC-1, Littlefuse 31201.5HXP 12-42VDC systems: 3 Amp, 250V, I2t, 14 A2s Rating, Fast Acting Bel Fuse 3AG 3-R, Littelfuse 312003P, Schurter 0034.5135 |
Line power fuses | 15W maximum – DC 40VA maximum – AC |
Switch-on current | AC: Maximum 35.7A (< 5ms) at 250VAC DC: Maximum 42A (< 5ms) at 42VDC |